ATI A15/66 Residual Sulfite Monitor
11
O&M Manual
Rev-W (8/10)
CHEMISTRY MODULE
The chemistry module is the heart of the sulfite monitor. Inside this module are the pumps, air
supply, air control instruments, and sensor element required to make an accurate sulfite measurement.
All items are housed in an ABS enclosure, with the main components mounted on the front panel inside
the clear cover. The enclosure is hinged in two places, with the front cover able to swing in either
direction, and the center compartment allowed to swing open on the right hand hinge (open left release
latch). Opening the center compartment allows access to the terminal circuit board, air pump, and
sample and reagent pump wiring.
Sulfite measurement in the chemistry module is done by pumping a small amount of sample out
of the inlet overflow chamber, mixing the sample with sulfuric acid to convert sulfite to sulfur dioxide gas,
and then air stripping the SO
2
for measurement in the gas phase. Figure 7 provides a view of the
chemistry module components.
Figure 7 - Chemistry Module Layout