ATI A15/66 Residual Sulfite Monitor
36
O&M Manual
Rev-W (8/10)
MONITOR RESET
The electronic unit can be reset to factory defaults if you wish to quickly get the monitor back to
its original operating configurations. To do this, go to step 1 of the configuration menu (page 23). Use
the UP arrow to change to H2S. Then press ENTER. Press mode once and then go back into step one
of the configuration menu and press the DOWN key to return to SO3. Press ENTER and the unit will be
reset to the factory defaults for sulfite. You can then continue through the configuration menu if you need
to reset any range, output, alarm, or other configurations.