ATI A15/66 Residual Sulfite Monitor
21
O&M Manual
Rev-W (8/10)
9. TEMPERATURE: ON Factory setting. No adjustment should be made. Press ENTER to go to the
next display.
10. CAL TC: 0.00%/C Factory setting. No adjustment should be made. Press ENTER to go to the
next display.
11. CAL OUT1: 4-20 mA This routine allows programming of the analog output signal. Either a 4-20 or
0-20 mA output may be selected by using the UP key. The default is 4-20 mA. If a zero based
voltage output is required, the 0-20 mA output may be used in conjunction with a shunt resistor
across the connection terminals on the back of the instrument. Once a selection is made, press
ENTER.
12. CAL P1: 0.00 This routine allows selection of the low end of the analog output range. The default
is 0.00 and should normally be left unchanged. However, the output can be adjusted to a non-zero
based range (such as 1-5 PPM) if desired. If a non-zero based output is desired, use the UP key to
set the new 4 mA output value. When selection is made, press ENTER.
13. CAL P2: 20.00 This routine allows selection of the full scale analog output range. The default is
20.00 PPM (or 2.000 PPM if the low range is selected in #3 above). Use the UP or DOWN keys to
set the concentration to be represented by a 20 mA output value. When selection is made, press
ENTER.
14. SET A ACT: ON/OFF This routine allows selection of the mode of operation for relay A. The default
is simple on/off operation. This relay can also be configured for pulse frequency (PFM) or pulse width
(PWM) modulation for control of chemical feed equipment.
15. SET A F.: LO (OR HI) This routine allows the function of setpoint A to be defined as either low or
high. When LO is selected, relay A acts like a low alarm, with the relay activating when the measured
concentration is below the setpoint. When HI is selected, the relay acts like a high alarm, activating
when sulfite is above the setpoint. Setpoints are programmed in another routine described later in
this manual.
16. SET B ACT: ON/OFF This routine allows selection of the mode of operation for relay B. The default
is simple on/off operation. This relay can also be configured for pulse frequency (PFM) or pulse width
(PWM) modulation for control of chemical feed equipment.
17. SET B F.: LO (OR HI) This routine allows the function of setpoint B to be defined as either low or
high similar to relay A described in 13 above.
18. CAL CF: DISABLED This function allows the user to activate an automatic cleaning fuction in the
monitor. The cleaning function is explained in greater detail later in this manual. Use the UP or
DOWN keys to select AUTO or MANUAL modes of cleaning. Leave this setting at DISABLED if no
automated cleaning is anticipated.
19. CLEANING T: 3.0’ This routine allows adjustment of the time period in which cleaning chemicals
can be injected into the sampling system. The default is 3 minutes and the range of adjustment is
from 0.1 to 20 minutes.
20. INHIBIT T: 4.0’ During the cleaning cycle specified above, the air pump is inhibited so that chlorine
from cleaning chemicals is not sparged across the sensor. The inhibit time should normally be
between 1.0 and 2.0 minutes longer than the cleaning time to allow cleaning chemicals to clear out of
the sample chamber before the sample pump is reactivated. The default is 4 minutes and the range
of adjustment is from 0.1 to 20 minutes.