ATI A15/66 Residual Sulfite Monitor
15
O&M Manual
Rev-W (8/10)
ACID PICKUP TUBING CONNECTION
A special rubber stopper with a ceramic support tube is supplied for the acid bottle. After the acid
bottle has been filled, insert the stopper into the bottle and press in securely. A length of pickup tubing is
also supplied for the stopper assembly. One end of this tubing should extend to the bottom of the acid
bottle inside the ceramic tube.
After the acid bottle has been placed into the bracket or on the mounting panel shelf, attach the
other end of the pickup tubing to the connector on the reagent pump. The reagent pump has a small barb
connector on the right side. Press the pickup tube carefully onto each of this barb, being careful not to
pull on the pump tubing. The tubing from the acid bottle can be shortened if there is an excess once
mounting is complete. Once all connections are made, be sure that the pickup tube extends to the bottom
of the bottle inside the ceramic guide.
Figure 9 - Solution Feed Assembly
CAUTION: Use extreme care in handling of the acid used with this analyzer.
Always wear eye protection and acid resistant gloves when working
with this acid. Sulfuric acid can cause eye damage and/or severe
burning of exposed skin.